- | ===== When to reconfigure the scanner | + | ===== When is it necessary? |
The LMI 3D scanner has three configurations: | The LMI 3D scanner has three configurations: | ||
- for small objects | - for small objects | ||
Line 18: | Line 18: | ||
===== Warnings ===== | ===== Warnings ===== | ||
- | | + | <WRAP center round important 60%> |
+ | * **Never use the menu of the projector**. | ||
* Be careful with the 3D scanner. | * Be careful with the 3D scanner. | ||
* Do not skip steps or you will have to start over. | * Do not skip steps or you will have to start over. | ||
+ | </ | ||
+ | |||
+ | | ||
===== 1. Hardware configuration ===== | ===== 1. Hardware configuration ===== | ||
+ | |||
+ | Set the camera configuration (close, mid, far). | ||
{{: | {{: | ||
Line 75: | Line 81: | ||
{{: | {{: | ||
{{: | {{: | ||
- | Set the projector focus. | + | Set the projector focus. |
===== 4. Focus cameras ===== | ===== 4. Focus cameras ===== | ||
Line 92: | Line 98: | ||
===== 5. Set exposure ===== | ===== 5. Set exposure ===== | ||
{{: | {{: | ||
- | Expose the image of the cameras well by changing the projector brightness. | + | Expose the image of the cameras well by changing the projector brightness. |
+ | Try to minimise under and over exposure on the calibration card. | ||
===== 6. Calibrating ===== | ===== 6. Calibrating ===== | ||
{{: | {{: | ||
Match the size of the squares of the calibration card.\\ | Match the size of the squares of the calibration card.\\ | ||
- | Make sure sensible Filter values are set. | + | Make sure sensible Filter values are set. (no documentation found on where the values are measured from. Possibly the CCD location of the camera' |
- | Delete | + | {{: |
+ | Click "Reset Calibration" | ||
{{: | {{: | ||
Capture shots of the calibration board. Move the calibration board to a different place for the next capture. Stay within the camera image as this will result in an invalid capture. It's not necessary to overlap the board with previous captures. Take your time and be patient. | Capture shots of the calibration board. Move the calibration board to a different place for the next capture. Stay within the camera image as this will result in an invalid capture. It's not necessary to overlap the board with previous captures. Take your time and be patient. | ||
- | {{: | + | {{: |
Try the following: | Try the following: | ||
- | * capture a horizontal row where the board faces the cameras | + | * capture a horizontal row where the board faces the cameras:\\ |
- | * do the same row, but now with the board at an angle | + | {{: |
+ | * do the same row, but now with the board at an angle:\\ | ||
+ | {{: | ||
* move the board further or closer to the camera for about 6cm and repeat the above steps | * move the board further or closer to the camera for about 6cm and repeat the above steps | ||
* do this for approximately the same depth as the scanning volume is wide. | * do this for approximately the same depth as the scanning volume is wide. | ||
Line 114: | Line 125: | ||
- | You will need 40 or more images for a qualitative | + | You will need 40 or more images for a good quality |
{{: | {{: |