lmi_3d_scanner:configuring_scan_volume

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
lmi_3d_scanner:configuring_scan_volume [2019/10/14 07:25] – [6. Calibrating] formlablmi_3d_scanner:configuring_scan_volume [2021/01/14 02:58] (current) – [6. Calibrating] formlab
Line 81: Line 81:
 {{:lmi_3d_scanner:projector_focus_wiki.jpg?nolink|}}\\ {{:lmi_3d_scanner:projector_focus_wiki.jpg?nolink|}}\\
 {{:lmi_3d_scanner:projector_focus_result.jpg?nolink|}}\\ {{:lmi_3d_scanner:projector_focus_result.jpg?nolink|}}\\
-Set the projector focus.+Set the projector focus. (Look at the actual calibration card, not the images from the cameras)
  
 ===== 4. Focus cameras ===== ===== 4. Focus cameras =====
Line 104: Line 104:
 {{:lmi_3d_scanner:calibration_board_square_size_and_filter_settings_wiki.jpg?nolink|}}\\ {{:lmi_3d_scanner:calibration_board_square_size_and_filter_settings_wiki.jpg?nolink|}}\\
 Match the size of the squares of the calibration card.\\ Match the size of the squares of the calibration card.\\
-Make sure sensible Filter values are set.+Make sure sensible Filter values are set. (no documentation found on where the values are measured from. Possibly the CCD location of the camera's).
  
  
Line 125: Line 125:
  
  
-You will need 40 or more images for a qualitative calibration.+You will need 40 or more images for a good quality calibration.
  
 {{:lmi_3d_scanner:calibrate-button.jpg?nolink|}}\\ {{:lmi_3d_scanner:calibrate-button.jpg?nolink|}}\\